AQUACOUNTER AQ-2100 Series Karl Fischer Coulometric Titrator |
 |
 |
AQUACOUNTER AQ-2100 Karl Fischer
Coulometric titrator is ideal for the determination of trace level
of water. The wide detection range from 5 ppm to 10%, with its maximum
capability for automation and future up-gradability, AQ-2100 leaves
nothing unsatisfactory. |
 |
Features |
 |
- A Large Size Color LCD Screen.
A large size Color LCD screen provides all information
which operator will need. Real-time titration curve can
be a great source for method development.
- Two-Channel expansion capability.
An optional coulometric kit is available that will convert
the unit to a two-channel coulometric titrator. With this
factory-installed kit, you'll be able to quickly and efficiently
run both routine samples and those problematic samples
that require special reagents (e.g., aldehydes and ketones).
- Innovative cell design decreases
maintenance over the life of the cell.
Replaceable membrane combines the superior precision of
a ceramic frit with the ease-of-maintenance of a fritless
system. Convenient, quick, and safe electrode cable attachment.
- GLP compliant instrument validation promises reliable results.
GLP compliant instrument validation functions are provided to promise reliable results, such as information of life of reagents, life of electrodes, etc. Complete control from a PC or laptop is now standard with RS-232 cable included.
|
|
 |
 |
Standard Specifications |

|
| Item |
Specifications |
| Titration method |
Coulometric Karl Fischer titration method |
| Electrolytic control method |
Constant current electrolysis, intermittent electrolysis in the vicinity of the end point only |
| Titration control method |
AC polarization potential difference detection method (1) Wait-time (INTERVAL) controlled; duration from 1 - 99 sec (2) Titration-time (T-TIMER) controlled; duration from 1 second to 99 minutes |
| Automatic background compensation range |
0 - 50 µg H2O/min |
| Measuring range |
10µg H2O - 99mg H2O |
| Accuracy |
Within CV of 0.3% With water-methanol (1mg H2O/mL),1g is measured 10 times. |
| Required measuring time |
FAST; 33 sec/mg H2O, MEDIUM; 50 sec/mg H2O,SLOW; 100 sec/mg H2O |
| Titration cell capacity |
150ml |
| Display screen |
Color graphic liquid crystal display |
| Display unit |
µg H2O (the least amount of display is 0.1µg H2O), %, and ppm |
| Calculation functions |
Concentration and statistics, Recalculation |
| Method files |
Titration parameter files; 20, Sample file; 1 (number of reserved samples = 99), EV file; 20 |
| Memory data capacity |
50 data points/channel |
| Printer interface |
One parallel port |
| External input / output |
Two RS-232C ports Port 1 for PC connection. Port 2 for Balance connection. |
| Power supply, Power consumption |
AC100 - 240V, 50/60Hz, 100 VA |
| Dimensions / Weight |
220 (W) x 400 (D) x 200 (H) mm
Approx. 5.0 kg |
| Safty approval |
CE approved |
|
|
 |
 |


















 

|