AQUACOUNTER AQ-2100 Series Options |
 |
 |
1. Solid Evaporator EV-2000 |
| |
 |
EV-2000 is designed for solid sample which is not soluble in Karl Fischer reagent or contains interfering substances. EV-2000 works with our AQUACOUNTER Karl Fischer titrator series and automatically controlled from AQ-2100 or AQV-2100 titrator. |
|
| |
EV-2000 Standard Specifications |
|

|
| Heating Temperature |
Room temperature to 350 |
| Carrier gas |
Nitrogen gas or dry air |
| Carrier gas flow rate |
30 - 300mL/min |
| Carrier gas input pressure |
0.8MPa or less |
| Carrier gas desiccant |
Molecular sieves |
| Heater |
Aluminium block heater |
| Sample volume |
About 10mL or less |
| Power supply, Power consumption |
AC100V - 240V, 50/60Hz, 150VA |
| Dimensions, Weight |
220 (W) x 430 (D) x 200 (H) mm, Approx. 8kg |
|
|
 |
 |
2. Oil Evaporator EV-2000L |
| |
 |
The EV-2000L Oil Evaporator utilizes azeotropic distillation (Micro Dean & Stark Distillation).It completely separates water from sample matrices, thus eliminates side reaction with interfering substances. EV-2000L can be used with AQUACOUNTER Karl Fischer titrators, or any other manufacturer's titrators. |
|
| |
EV-2000L Standard Specifications |
|

|
| Evaporation method |
Azeotropic distillation method |
| Heating range |
Room temperature to 200 |
| Carrier gas |
Nitrogen gas or dry air |
| Carrier gas flow rate |
30 - 300 mL/min |
| Carrier gas inlet pressure |
0.8 MPa or less |
| Carrier gas desiccant |
Molecular Sieves |
| Heater, Heating block material |
Aluminium block heater |
| Sample volume |
About 10ml or less |
| Measurement range |
The same as that of the connected titrator |
| Azeotropic Solvent |
Benzene, Toluene, Xylene, etc. Volume for one measurement; 5-10mL. |
| Power supply, Power consumption |
AC 100V - 240V, 50/60Hz, 150 VA. |
| Dimensions, Weight |
220 (W) x 370 (D) x 265 (H) mm, Approx. 7kg |
|
 |
 |
3. Automated Solid Evaporator EV-2010 |
| |
 |
The EV-2010 automated evaporator/sample changer can be connected with both AQ-2100 coulometric titrator and AQV-2100 volumetric titrator, which makes AQS-2110 a complete automated evaporation KF titration system. |
|
| |
- For medium to heavy workloads
Up to 10 samples can be mounted on the turntable at the same time. Up to 99 samples can be run by simply removing the finished sample and adding a new one.
- Extremely compact design
A compact upright design for both the heating block and the KF titrator saves your valuable bench space.
- Wide temperature range and programmable temperature control
The heating temperature can be set up to 300
and programmed up to 3 steps. This feature makes it easier to find the optimum temperature for each individual sample, or allows you to differentiate surface water from bound water.
- Dry Air Pump (Optional)
Compact, integrated air pump eliminates the need for nitrogen carrier gas.
|
|
| |
EV-2010 Standard Specifications |
|

|
| Number of sample vials |
10 |
| Volume of sample vial |
10mL |
| Seal |
PTFE cap with aluminium foil |
| Carrier gas |
Nitrogen or Dry air |
| Carrier gas flow rate |
100 - 300mL/min |
| Carrier gas desiccant |
Molecular Sieves |
| Heater, Heating block material |
Aluminium block heater |
| Heating range |
Room temperature to 300 |
| Power supply, Power consumption |
AC100 - 240V, 50/60Hz, 150VA |
| Dimensions, Weight |
260 (W) x 600 (D) x 280 (H) mm, Approx. 11kg |
|
|
 |
 |
4. Special Sampling Devices |

|
| Name |
Description |
Figure |
| Micro Solid Sampler |
For Solid/powder sample. Sample size up to 3mL |
 |
| Micro - Micro Solid Sampler |
For solid/powder sample. Especially for very small amount of sample.(sample size up to 0.17mL) |
 |
| Grease Sampler |
For viscous samples such as grease |
 |
| Gas sample foaming tube |
A device for introducing a gas sample into A titration cell. |
 |
| Liquefied Gas Sampler |
For liquefied gas sample. |
 |
| Sampler |
A container for powder or pellets. Use with solid evaporator EV-2000. |
 |
|
|
 |
 |
5. Ohter Option |

|
| Model Name |
Description |
| PR-2000T2 |
Thermal Printer w/cable |
| PR-302B |
Dot Impact Printer w/cable |
| Printer Bracket 2000A |
To set up PR-2000T2 on AQ-2100 series |
| Printer Bracket 302A |
To set up PR-302B on AQ-2100 -series |
| K-2000A2 |
Magnetic Stirrer |
| K-2000A2P |
Magnetic Stirrer with pump for solvent add/drain |
|
 |
| |
Note: When ordering printer, please specify power voltage, i.e. AC100-120V or AC220-240V. |
|
 |




















|