| Item |
AQV-2100 |
| Titration method |
Volumetric Karl Fischer reagent titration method |
| Titration control method |
Intermittent titrant addition while approaching the end point. (1) Wait-time (INTERVAL) controlled; duration from 1 - 99 sec (2) Titration-time (T-TIMER) controlled; duration from 1 second to 99 minutes |
| Background correction |
Automatic correction can be enabled/disabled. |
| Buret |
Capacity; 10mL, reproducibility; ±0.01mL(for full discharge, standard deviation for n=5) Accuracy: within ±0.15% (for full discharge ,n=5) |
| Liquid contact material |
PTFE, ceramic and glass. |
| Valve selection |
Automatic selection with ceramic joint valve |
| Number of burets |
1; (expandable to 2) |
| Measurement range |
0.1mg H2O - 500mg H2O |
| Minimum volume of titrant addition |
0.005mL or 0.01mL |
| Titration cell capacity |
150mL, minimum volume of consumption; about 30mL |
| Display screen |
Color graphic liquid crystal display |
| Display units |
mL (min. display is 0.001mL.), mg H2O, %, and ppm |
| Calculation functions |
Concentration, factor standardization, and statistics,and Recalulation |
| Method files |
Titration parameter files; 20, Sample file; 1 (number of reserved samples = 99), EV file; 20 |
| Memory data capacity |
50 data points/channel |
| Printer interface |
One parallel port |
| External input/ output |
Two RS-232C ports Port 1 for PC connection Port 2 for Balance connection |
| Power supply, Power consumption |
AC100 - 240 V, 50/60Hz, 100 VA |
| Dimensions |
220 (W) x 400 (D) x 330 (H) mm |
| Weight |
Approx. 7kg (Stirrer and printer not included.) |