Accurate measurement of spectra for Fluorescence Spectrophotometer


  A decrease in irradiated light is measured by absorption spectroscopy, while luminescence is observed by the fluorescence method; therefore, the wavelength differs between irradiated and measured light.
The fluorescence method has the advantage of being more sensitive (measurable at a low concentration) and more informative (acquisition of excitation and fluorescence spectra), as compared with the absorption spectroscopy.
 
  However, wavelength characteristics specific to optical components of a fluorophotometer, such as a lamp, may cause differences among devices, and multiple lights such as secondary light, due to the effects of light scattering, may disturb measurement. Here, this is to introduce you a method for correcting (reducing) the differences among devices and a method for eliminating light scattering.  

 Effect to use Filter set
 Effect to use Spectra Correction Kit

Effect to use Filter set


  In surface photometry, multiple lights such as secondary light, due to the effects of light scattering, disturb measurement.
A cut filter is effective for eliminating multiple lights.
We utilize the Stoke's principle that fluorescence can be measured at a longer wavelength, as compared with excitation light.
Filter Set  
Filter Set
  We'll introduce the effects of the cut filter using data obtained for sodium salicylate (powder). Use of a filter (UV35) cuts light scattering, led to a detector, to control the generation of secondary light.  

  Effect to use Filter set  
  This is data extracted from three-dimensional data. From two dimensional data, the 2nd light is eliminating. 2nd date uses Filter set  

Effect to use Spectra Correction Kit


  This is to introduce you the spectrum correction to get the accurate spectrum. Each lamp or detector of a fluorophotometer has its own wavelength characteristics.
Since the measured spectrum data reflects such characteristics, correction is needed to obtain a true spectrum. Here, we show the effects of a spectrum correction kit that generally includes a rhodamine B concentrated solution and a substandard light source.
Spectra Correction Kit  
Spectra Correction Kit
  These are the effects after spectrum correction using data obtained for sodium salicylate (powder). Three-dimensional data varied with the presence or absence of correction, demonstrating the effects of wavelength characteristics.  

  Effect to use Spectra Correction Kit  

  2nd data uses spectra correction Kit  
  This is a comparison between literature data and measurement data. We compared certificated values in the presence or absence of spectrum correction for NIST SRM 936a quinine sulfate. Comparing spectra in the literature or between other devices also demonstrates effectiveness of the correction.  

Related Products
F-7000 Fluorescence Spectrophotometer  F-7000 Fluorescence Spectrophotometer
F-2700 Fluorescence Spectrophotometer  F-2700 Fluorescence Spectrophotometer

 

Search by Google
 > Advanced Search
Life science

Topics

Global Network

Products

starting of secondary navigation UV-Visible/NIR Spectrophotometer

Fluorescence Spectrophotometer

starting of secondary navigation F-7000

starting of secondary navigation Atomic Absorption Spectrophotometer

starting of secondary navigation HPLC/Amino Acid Analyzer

starting of secondary navigation LCMS

starting of secondary navigation Clinical Analyzer

starting of secondary navigation Potentiometric Titrator

starting of secondary navigation Karl Fischer Titrator

starting of secondary navigation Mercury Analyzer

Applications

Contact Us

Sitemap

pagetop
 
starting of footer  | Terms of use | Privacy Policy | All Rights Reserved Copyright (C) 2001, 2012. Hitachi High-Technologies Corporation.
starting of footer  | Terms of use | Privacy Policy | All Rights Reserved Copyright (C) 2001, 2012. Hitachi High-Technologies Corporation.