Performance (results of measuring 5 samples by protein hydrolyzate assay (standard) method) |
| Analysis time |
Approx. 30 minutes (net) |
| Reproducibility of peak retention time |
CV0.3% (Arg), 0.5% (Ala) |
| Reproducibility of peak area |
CV1.0%(Gly, His) |
| Detection limit |
3 pmol (S/N = 2, Asp) |
| Analyzer |
| Column |
4.6 mm ID ' 60 mm L*1 |
| Packed with Hitachi custom ion exchange resin |
| Pump |
Delivery pressure; 0 to 30 MPa (0 to 306 kgf/cm2) |
| Flow rate range; 0.000 to 0.999 mL/min |
| Autosampler |
Injection methods; Direct injection method |
| Sample vial capacity; 1500 µL |
| No. of vials accommodated; 200 (200 when optional cooler
installed) |
| Sample injection volume; 1 to 100 µL |
| Column oven |
System; Peltier |
Temperature setting range; 20 to 85
(in 1 steps) |
| Reaction unit |
System; Electronic heating (reaction column) |
Temperature setting; 50 to 140 (in 1 steps) |
| Photometer |
Spectrophotometer; Aplanatic concave diffraction grating |
| Wavelength; 570 nm, 440 nm Analyzer |
| EZChrom Elite for HITACHI Amino Acid Analyzer |
| CPU |
32 bit OS, Windows XP®*2 |
| CRT |
43.2 cm color display |
| Dimensions (excluding PC) |
900(W) ' 650(D) ' 1050(H) mm |
| Weight (excluding PC) |
Approx. 250 kg |
| Operating temperature range |
15 to 35 |
| Power supply |
100 115
V AC / 220 240 V AC, 800 VA and over, w/in 50/60 Hz ±0.5 Hz |
| N2 gas |
N2 gas source must be prepared. |