UV-Vis/NIR Spectrophotometer Basic Course
7. Structure of a spectrophotometer

UV-Vis/NIR Spectrophotometer Basic Course
 UV-Vis/NIR Spectrophotometer Basic Course

Optical system


  Is there a dedicated position for a sample? (Single beam and double beam)  
  Single beam system
In the measurement of absorption, a sample and a reference are measured each time to compare these measurement results.
Since a single beam system has only one position for measurement, the sample and the reference should be replaced, and 100% adjustment (or 0 Abs adjustment) should be conducted whenever measurement wavelength is changed.
Although this system is used mainly for quantitative measurement, and offers a small and inexpensive device configuration, changes (drift) in the light source cannot be adjusted.
 
  Single beam system  
 
 
  Double beam system
A double beam system has two dedicated positions for the sample and the reference, in contrast to the single beam system with only one measurement position. The double beam system measures sample based on this signal because the light of reference enters into Detector.
The change of the light source can be corrected, and it is possible to measure the system with stability.
 
  Double beam system  


  Careful selection of wavelength (single and double monochromes)  
  A monochrome (monochromator, spectroscope) plays the role of a sorter that picks up the light of the desired wavelength out of the light containing various wavelengths.
But, however carefully they are checked, wavelengths other than the wavelength of interest may slip in (this light that was lost or slipped in is called stray light).
This stray light may affect the accuracy of the measurement. A double monochrome system conducts a second check on the light that passed the first check. The system that checks only once is called a single monochrome system.
 
  Single monochrome & Duble monochrome  

 The next chapter is "Components of spectrophotometer".
Previous Next

Related Products
U-4100  U-4100 UV-Visible-NIR Spectrophotometer
U-3900/3900H  U-3900/3900H Spectrophotometer
U-2900/2910  U-2900/2910 Double Beam Spectrophotometer
U-5100 Ratio Beam Spectrophotometer  U-5100 Ratio Beam Spectrophotometer
Diffraction Grating  Diffraction Grating
contents end
Search by Google
 > Advanced Search
Life science

Topics

Global Network

Products

UV-Visible/NIR Spectrophotometer

starting of secondary navigation U-4100

starting of secondary navigation U-3900/3900H

starting of secondary navigation U-2900/2910

starting of secondary navigation U-5100

starting of secondary navigation Fluorescence Spectrophotometer

starting of secondary navigation Atomic Absorption Spectrophotometer

starting of secondary navigation HPLC/Amino Acid Analyzer

starting of secondary navigation LCMS

starting of secondary navigation Clinical Analyzer

starting of secondary navigation Potentiometric Titrator

starting of secondary navigation Karl Fischer Titrator

starting of secondary navigation Mercury Analyzer

Applications

Contact Us

Sitemap


pagetop
 
starting of footer  | Terms of use | Privacy Policy | All Rights Reserved Copyright (C) 2001, 2012. Hitachi High-Technologies Corporation.
starting of footer  | Terms of use | Privacy Policy | All Rights Reserved Copyright (C) 2001, 2012. Hitachi High-Technologies Corporation.