UV-Vis/NIR Spectrophotometer Basic Course
8. Structure of a spectrophotometer

UV-Vis/NIR Spectrophotometer Basic Course
 UV-Vis/NIR Spectrophotometer Basic Course

Components of spectrophotometer


  Light source  
  Two kinds of lamps, a Deuterium for measurement in the ultraviolet range and a tungsten lamp for measurement in the visible and near-infrared ranges, are used as the light sources of a spectrophotometer.  
 
 
  Types and properties of light source  
 
Type Tungsten Deuterium
Sign W WI D2
Property A continuous spectrum of 300-3000nm is emitted. A continuous spectrum of 168-500nm, with maximum energy at 250nm, is emitted.
Wavelength range 340-1100nm 185-360nm
Spectrum energy Spectrum energy of tungsten Spectrum energy of deuterium
 


  Spectroscope  
  A spectroscope plays a role in selecting a monochromatic light from a light source (white light).
Spectroscopes include Filter type, Prism type, and Grating (diffraction grating) type.
 
 
 
  Types and properties of dispersion  
 
Type Filter Prism Grating
(diffraction grating)
Properties A single wavelength can be extracted with a filter. A filter is also used in combination with diffraction grating for filtering out stray light. A spectrum of 175-2700 nm can be dispersed. The degree of dispersion varies with the wavelength. Dispersion is homogeneous at the entire wavelength, and a wide-range wavelength can be obtained with a diffraction grating. In addition, a constant spectrum featuring a constant slit breadth can be obtained.
Types and materials
Colored glass filter Interference filter Crystal or fused quartz Plane diffraction grating Concave diffraction grating
 


  Sample cell  
  A container that contains a sample is usually called "cell"; two types are available, glass and quartz cells. Since light in the ultraviolet range with a wavelength of 340 nm or less hardly passes through a glass cell, it is used for measurement in the visible range of 340 nm or more.
On the other hand, although a quartz cell allows passage of light in the entire wavelength in the ultraviolet and visible ranges, it is mainly used for the measurement in the ultraviolet range due to its high price.
 
  Sample cell  


  Detector  
  A detector plays a role in changing the light transmitted from a sample into an electric signal.
Optical semiconductor, Types of photomultiplier, etc. are available.
 
 
 
 
Optical semiconductor Photomultiplier
High speed, high sensitivity, and a low noise are features over the wavelength from Ultraviolet to Near infrared. A photo cell sensitive only in the visible range is used. A Si photodiode is a representative one.
Optical semiconductor
This combines a photoelectric tube and an amplifier (approximately 10-fold amplification) sensitive in both the ultraviolet and visible regions, and the sensitivity can be changed sharply by increasing the applied voltage.
Photomultiplier
 

 Finally, we introduce "UV-Visible/NIR Spectrophotometer of Hitachi".
Previous

Related Products
U-4100  U-4100 UV-Visible-NIR Spectrophotometer
U-3900/3900H  U-3900/3900H Spectrophotometer
U-2900/2910  U-2900/2910 Double Beam Spectrophotometer
U-1900  U-1900 Ratio Beam Spectrophotometer
U-5100 Ratio Beam Spectrophotometer  U-5100 Ratio Beam Spectrophotometer
Diffraction Grating  Diffraction Grating

contents end
Search by Google
 > Advanced Search
Life science

Topics

Global Network

Products

UV-Visible/NIR Spectrophotometer

starting of secondary navigation U-4100

starting of secondary navigation U-3900/3900H

starting of secondary navigation U-2900/2910

starting of secondary navigation U-1900

starting of secondary navigation U-5100

starting of secondary navigation Fluorescence Spectrophotometer

starting of secondary navigation Atomic Absorption Spectrophotometer

starting of secondary navigation HPLC/Amino Acid Analyzer

starting of secondary navigation LCMS

starting of secondary navigation Clinical Analyzer

starting of secondary navigation Potentiometric Titrator

starting of secondary navigation Karl Fischer Titrator

starting of secondary navigation Mercury Analyzer

Applications

Contact Us

Sitemap


pagetop
 
starting of footer  | Terms of use | Privacy Policy | All Rights Reserved Copyright (C) 2001, 2010. Hitachi High-Technologies Corporation.
starting of footer  | Terms of use | Privacy Policy | All Rights Reserved Copyright (C) 2001, 2010. Hitachi High-Technologies Corporation.